DNA shearing with the Bioruptor®
The best parameters to shear your DNA 150bp to 1kb are a few clicks away. These optimal settings have been validated by our team and can guide you with your DNA shearing experiments. Diagenode recommends that the concentration of DNA should not exceed 50ng/ul for best shearing reproducibility.
find your settings
 
Which Bioruptor® do you use?

Achieving proper fragment sizes for input DNA is crucial for your experiments and Diagenode has developed the most advanced water bath sonicators, the Bioruptor®. The Bioruptor® Plus is specifically designed for ChIP-PCR and it can also be used for DNA methylation analysis. The Bioruptor® Pico (modern evolution of the Bioruptor® NGS) is the most advance system for NGS. The Diagenode One is the new desktop solution that provides optimal DNA shearing.




 
Which size do you need your sheared DNA?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Which volume do you shear?


 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 30 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.65ml tubes (C30010011) and the corresponding 0.65ml tube holder and shear your samples for 30 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 20 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 13 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.65ml tubes (C30010011) and the corresponding 0.65ml tube holder and shear your samples for 13 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 15 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 13 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 7 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.65ml tubes (C30010011) and the corresponding 0.65ml tube holder and shear your samples for 6 cycles (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 8 cycles (sonication parameters: 15''ON and 90''OFF).

 
Use the Diagenode 0.65ml tubes (C30010011) and the corresponding 0.65ml tube holder and shear your samples for 8 cycles (sonication parameters: 15''ON and 90''OFF).

 
Use the Diagenode 1.5ml tubes (C30010016) and the corresponding 1.5ml tube holder and shear your samples for 6 cycles (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 7 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 1.5ml tubes (C30010016) and the corresponding 1.5ml tube holder and shear your samples for 20 cycles (sonication parameters: 15''ON and 15''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 6 cycles (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.1ml tubes (C30010015) and the corresponding 0.1ml tube holder and shear your samples for 6 cycles (sonication parameters: 5''ON and 5''OFF).

 
Use the Diagenode 0.65ml tubes (C30010011) and the corresponding 0.65ml tube holder and shear your samples for 7 cycles (sonication parameters: 5''ON and 5''OFF).

 
Which size do you need your sheared DNA?


 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes  and the corresponding  tube holder and shear your samples for 70 cycles or 70min (sonication parameters: 30''ON and 30''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 30 cycles or 60min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 10 cycles or 20min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 8 cycles or 16min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 6 cycles or 12min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 5 cycles or 10min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 3 cycles or 6min (sonication parameters: 30''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 4 cycles or 7min (sonication parameters: 15''ON and 90''OFF).

 
Use the Diagenode 0.5ml (C30010013) or 1.5ml (C30010010) TPX tubes and the corresponding  tube holder and shear your samples for 2 cycles or 3.5min (sonication parameters: 15''ON and 90''OFF).

 
Diagenode One gives fragment size of 200 bp after 90 s of sonication* (human genomic DNA at 10 ng/µl).  We recommend that you start with a time-course experiment. Our experts are available to assist you at https://diagenode.freshdesk.com/support/home .

* The Diagenode One operates through sonication and cooling cycles. The perfect control of the temperature provides optimal DNA shearing and sample quality protection. 90 s of sonication means a total run time of about 10 min.
 
Which Diagenode One Microfluidic chips do you use?


 
Diagenode One 20 µl Microfluidic chips

The Diagenode One operates through sonication and cooling cycles. We recommend that you select the best Sonication time to suit your needs. The Total time gives you an idea of the expected time for one run. Our experts are available to assist you at https://diagenode.freshdesk.com/support/home.
 
Diagenode One 50 µl Microfluidic chips

The Diagenode One operates through sonication and cooling cycles. We recommend that you select the best Sonication time to suit your needs. The Total time gives you an idea of the expected time for one run. Our experts are available to assist you at https://diagenode.freshdesk.com/support/home.
Thank you and we wish you a good
shearing experience with the Bioruptor®.